| Brand Name: | SKYLINE |
| Model Number: | SL-FL01 |
| MOQ: | 1 |
| Price: | Negociate |
| Supply Ability: | 1 Set/Month |
The SL-FL01 Cone Calorimeter is the single most important instrument for assessing the fire behavior of building materials, composites and finished products. Under controlled irradiance from a conical radiant heater, it measures heat release rate (HRR), total heat released, smoke release, time to ignition, oxygen consumption, carbon monoxide and carbon dioxide generation, and mass loss rate.
A specimen is mounted horizontally beneath a conical electric heater and exposed to a selectable radiant heat flux, typically 0–100 kW/m². Exhaust gases pass through a hood where oxygen concentration is measured by paramagnetic or zirconia analyzers. Heat release rate is calculated by oxygen-consumption calorimetry (approximately 13.1 MJ per kg of O₂ consumed). Smoke is measured by laser or photometric obscuration in the exhaust duct, and mass loss is tracked by a load cell. Results yield HRR, peak HRR, THR, FIGRA, SMOGRA and ignition time.
| Heat flux range | 0 – 100 kW/m² (adjustable conical heater) |
|---|---|
| Specimen size | 100 mm × 100 mm (standard), thickness up to 50 mm |
| Oxygen analyzer | Paramagnetic / zirconia, accuracy ±0.001 vol% O₂ |
| Exhaust gas flow | Controlled by mass-flow / blower system |
| Smoke measurement | Laser / photometric obscuration in exhaust duct |
| Load cell | Mass loss resolution 0.01 g |
| Ignition spark | Standard pilot burner with automatic spark |
| Ambient | 0.1 μg |
| Electricity | 0–450 mA automatic control |
| Gas supply | Propane (min. purity 95%), methane for calibration, N₂ / CO₂-CO mixtures |
| Dimensions | Approx. 1800 mm (W) × 900 mm (D) × 2650 mm (H) |
| Weight | Approx. 350 kg |
| Power supply | 220 V, 50 A (three-phase as required) |
| Ambient temperature | 10 °C to 35 °C |
It measures heat release rate via oxygen consumption, plus time to ignition, total heat released, smoke production, mass loss and gas generation. These are the primary inputs to fire safety classification (e.g., EN 13501, EN 45545).
ISO 5660-1, ASTM E1354 / D6113, ASTM E1740, ASTM F1550 and EN 45545-2 among others, making it the core apparatus for reaction-to-fire and smoke testing worldwide.
The heat released per unit of oxygen consumed is nearly constant (approximately 13.1 MJ/kg O₂) across fuels, so measuring the oxygen drop in the exhaust gives an accurate, fuel-independent heat release rate.
Smoke is recorded as specific extinction area (SEA) and smoke growth rate (SMOGRA) from light obscuration in the exhaust duct — key metrics for rail and marine material acceptance.