| Brand Name: | SKYLINE |
| Model Number: | SL-FL05 |
| MOQ: | 1 |
| Price: | Negociate |
| Supply Ability: | 1 Set/Month |
The SL-FL05 Radiant Flooring Panel Apparatus measures the critical radiant flux (CRF) of floor-covering materials. Critical radiant flux is the incident heat per unit area at which a material sustains flame spread; a higher CRF indicates a more stable material that is less likely to catch fire. The apparatus is equipped with a radiant panel and a T-type burner, and the farther the specimen is from the radiation source, the lower the flux it receives.
A horizontal floor-covering specimen is irradiated by an inclined radiant panel that produces a flux gradient along its length. A T-type pilot burner ignites the specimen at the high-flux end. The flame spreads until it self-extinguishes, and the distance from the burner to the extinction point corresponds, via the flux gradient, to the critical radiant flux. This value is used to classify flooring for fire safety.
| Maximum operating power | 5 kW |
|---|---|
| Timing accuracy | <1 s / h |
| Heat flux meter measuring range | (0–15) kW/m² (heat flow meter customer-owned) |
| Heat flux meter accuracy | ±0.2 kW/m² |
| Heat flux meter accuracy (repeatability) | <±3% |
| Heat radiation pyrometer range | 400–600 ℃, accuracy ±0.5 ℃ |
| Black body temperature error | ±5 ℃ |
| Smoke speed | (2.5 ± 0.2) m/s |
| Smoke capacity | (39–85) m³/min |
| Porous refractory radiation board size | 300 mm × 450 mm, capable of approximately 900 ℃ high temperature |
| Thermocouple | 3.2 mm diameter K-type stainless steel armored thermocouple |
| Burning torch | Ø10 mm and Ø6 mm diameter, spray tube with 35 × Ø0.7 mm holes |
| Flame height | 60–120 mm adjustable |
| Ignition system | High-voltage electronic ignition |
| Burning box size | L1420 × W1200 × H2300 (mm) |
| Strong exhaust system | The machine is equipped with a strong fan; when the test is completed, the exhaust fan system is started to exhaust outdoors |
CRF is the radiant heat flux (kW/m²) at the point where flame spread self-extinguishes. A higher CRF means the flooring resists fire spread better and achieves a better classification.
ISO 9239-1, ASTM E648, ASTM E970 and NFPA 253 all use the radiant panel; GB/T 11785 and GB 8624 are the Chinese equivalents.
The extinction distance is converted to flux using the calibrated gradient; that value is the CRF reported for classification.
Floor coverings are walked on and irradiated from above in fires; the horizontal orientation with a flux gradient realistically simulates that exposure.