1、The automatic micro-moisture tester should be placed in a dry place when not in use.
2、Don't put the instrument to work in the environment with corrosive gas.
3、Do not place the instrument in an environment with high humidity and large power fluctuations to avoid abnormal phenomena.
4、In case of poor sealing effect of the silica gel pad inside the inlet port, replace the silica gel pad according to the instruction.
5, a long time application plug, socket will adhere to the dirt, so that its poor contact, then the application of ethanol to wipe the metal parts respectively to make good contact.
6, the day after the test, be sure to empty the system of Karl Fischer reagent, and then cleaned with methanol, never use water to clean the system, because it is not easy to volatilize, will cause the next test Karl Fischer reagent calibration is not true.
7, automatic micro-moisture tester failure should immediately cut off the power, ask professionals to repair and troubleshoot before continuing to use, to prevent accidents
Contact Person: Ms. Ivy Xie
Tel: 86-13751491529
Fax: 86-769-38818154
Over Temperature Environmental Test Chamber With Temperature Humidity Stability Control
Iec 60529 Stainless Steel Envirotronics Chamber Free Dust Blasting Sand And Powder
3KW 150l Xenon Lamp Aging Test Chamber Air Cooled
Ip5/6x Programmable Constant Climate Chamber Power Sand And Dust Test
ISO 5660 AC220V Cone Calorimeter For Building Materials Testing
Fire Testing Equipment Steiner Tunnel Test Apparatus
Automatic Oil Moisture Analyzer Coulometric Karl Fischer Titrator
Full Automatic Closed Cup Flash Point Tester
0-80kv ASTM D877 ASTM D1816 Transformer Oil Breakdown Voltage Tester
Oil Analysis Testing Equipment Automatically Kinematic Viscosity Meter For Petroleum Product