    Product Details:

    Place of Origin: China
    Brand Name: SKYLINE
    Certification: CE
    Model Number: SL-E10

    Payment & Shipping Terms:

    Minimum Order Quantity: 1 unit
    Price: neogitable
    Packaging Details: plywood case
    Delivery Time: 10 work days
    Payment Terms: T/T, Western Union, MoneyGram
    Supply Ability: 1 unit / month
    Contact Now
    Detailed Product Description
    SKYLINE: Tracking Index Tester Standards: IEC60695,GB4207/IEC60112
    Electrode Dimensions: Width: 5mm±0.1mm Thickness: 2mm±0.1mm,
    Electrode Pressure: 1.00N±0.05N Power Supply: 800VA ,220V, 48-60Hz

    Flammability Test Apparatus , Tracking Index Tester For Evaluating Leakage Resistance

     

    This Tracking Index Apparatus is strictly designed and manufactured in accordance with IEC60112,IEC60335-1, IEC60598-1 and GB4207

    Measure the relative resistance to traceability of the surface of the solid electrical insulating material when exposed to impurities containing water at a voltage of up to 600 V.
    When the voltage is applied between the electrode means provided on the surface of the material and the electrolyte is dropped between the two electrodes at a predetermined time interval, it is possible to generate an electrical mark in this test case. The number of droplets necessary to cause material damage increases as the applied voltage decreases, and at less than a certain limit, no traceability occurs.
    When the material in the highest test voltage is not electric marks, can have different degrees of corrosion, and corrosion depth can be measured. Some materials can burn during the test.

     

    Technical Parameters

     

     Electrode Materials  The electrode caput of brass-platinum
     Electrode Distance  4.0mm±0.1mm,60±5°angle
     Test Voltage  100~600V adjustable, displayed value: r.m.s

     Test Current

     

     

     Current limiting 1A±0.1A adjustable,
       displayed value:r.m.s, tolerance of 1.5%

     Electrode Dimensions

     

     

     width: 5mm±0.1mm,
       Thickness: 2mm±0.1mm,
       length >12mm,
       end chisel-edge angle : 30°±2°,platinum.
     Electrode Pressure  1.00N±0.05N
     Dripping Nozzle  outer diameter: 0.9mm~1.2mm for solution A
       0.9mm~3.45mm for solution B
     Dripping Interval

     0-99 seconds (adjustable),the time for 44-55/1cm3(adjustable)

    drops to fall on the specimen shall be (30±2)min.

     Heigh Drops  30-40mm±5mm(adjustable)
     Control  Using single chip microcomputer + touch screen control
     Chamber volume  >0.5m³, black interior
     Outer Dimensions  W*D*H=1100mm*700mm*1300mm
     Power Supply  800VA ,220V, 48-60Hz

     

     