Horizontal Vertical Flammability Testing Equipment

Horizontal Vertical Flammability Testing Equipment

  • Horizontal Vertical Flammability Testing Equipment
  • Horizontal Vertical Flammability Testing Equipment
  • Horizontal Vertical Flammability Testing Equipment
Horizontal Vertical Flammability Testing Equipment
Product Details:
Place of Origin: China
Brand Name: SKYLINE
Certification: CE
Model Number: SL-S29
Payment & Shipping Terms:
Minimum Order Quantity: 1 unit
Price: neogitable
Packaging Details: Plywood case
Delivery Time: 10 working days
Payment Terms: T/T, Western Union, MoneyGram
Supply Ability: 1 unit / month
Contact Now
Detailed Product Description
SKYLINE: Vertical Flammability Tester Standards: UL 94-2006 IEC 60695-11-3 IEC 60695-11-4 GB/T5169.15 16 17 22-2008
Timer: 0-999.9s+/-0.1s (adjustable) Flame Height:: 20mm+/-2mm To 180mm+/-10mm(adjustable)
Weight: 128kg Gas: 98% Methane
High Light:

horizontal flammability tester

,

flammability test chamber

IEC60695-11-3  UL94 Horizontal-Vertical Flame Chamber
Product information
Horizontal and Vertical Flammability Testing Machine
UL 94 flame chamber. Horizontal flame chamber, vertical flame chamber
For the flammability testing of plastic materials used in consumer electric devices and appliances. Stainless steel construction with black interior and large sliding window for easy viewing. Includes a fully adjustable specimen support and precision gas flow controls.

 
Technical parameters

 Kinds 500W(5V class ) 50W(HB/Vclass)
 Test Flame 500W(burning test for V-0,V-1,V-2 classification) 50W(burning test for V-0,V-1,V-2 classification)
 Burner Tube Internal diameter: Φ 9.5mm±0.3mm Internal diameter: Φ 9.5mm±0.3mm,
 Burning Angle 0°,20°,45° 0°,20°,45°
 Thermocouple Type K (Ni/Cr - Ni/Al) Type K (Ni/Cr - Ni/Al)
 Thermocouple Size  0.5 mm 0.5 mm
 Standard Copper Block

 9mm±0.01mm,
weight:10g±005g before drilling.

 5.5mm±0.01mm,
weight:76g±0.01g before drilling.

 Time for Temperature to Increase From 100°C±5°C to 700°C±3°C; 54s±2s (preset)

 From 100°C±5°C to 700°C±3°C;
44s±2s (preset)

 Gas Flow 965±32ml/min 105±10ml/min
 Gas Supply methane,min.98% purity methane,min.98% purity
 Water-column 125m±5mm‹10mm
 Test Time 54±2s 44±2s
 Flame Height Indicator 40±1mm 20±1mm
 Volume Of Chamber >0.5m³, black interior >0.5m³, black interior
 Conform to standard UL94,IEC60695-11-3,IEC60695-11-20 UL94,IEC60695-11-4,IEC60695-11-10

 
Testing Standards
UL 94-2006 IEC 60695-11-3 IEC 60695-11-4 

UL94 Horizontal & Vertical Flammability Tester.pdf...
Horizontal Vertical Flammability Testing Equipment 0
 
Specimen holder
Horizontal Vertical Flammability Testing Equipment 1
 
Controll panel

 
How do you test for flammability?
The various flammability tests are all similar in that they (1) orient a test sample either horizontally or vertically and place it in the test chamber; (2) apply a flame from a Bunsen Burner for a specified time; and (3) measure the time or distance that the flame propagates.

Contact Details